5. Aspirate the lidocaine by using the 18 gauge Monoject Magellan
Safety Needle. Twist syringe clockwise while holding the rigid sheath
that covers the Monoject Magellan
Safety Needle to seat the
Safety Needle to the syringe. Remove the sheath from the Monoject
Magellan
Safety Needle while directing the needle away from you
and others. Be certain not to touch the needle or the safety shield.
6. If necessary to transport the filled syringe to the point of administration,
use a safe, passive recapping technique to cover the needle before
transport. In accordance with OSHA standards, such recapping must be
accomplished by a one-handed technique, i.e., do not hold the needle
shield during the recapping process.
7. Replace the needle used in Step 5 by following the safety precautions
mentioned in step 9.
8. Raise skin wheel with lidocaine by using the 25 gauge safety needle at
puncture site. The 22 gauge safety needle may be used for deeper
infiltration.
9. Immediately following injection procedure, lock the safety needle shield,
using any of the following methods:
a. Push the safety shield forward in a single-handed manner to cover the
needle and lock the shield. Keep your finger or thumb behind the
needle tip --at all times.
OR:
b. Press the safety shield, lever side down, onto a flat surface to
cover the needle and lock the shield, i.e., bedside table.
The Monoject Magellan
Safety Needle is locked once
the needle tip is completely shielded. Verify locked position
through audible, tactile, and/or visual feedback.
Only grasp the colored hub to connect/disconnect needle and
syringe.
Once safety shield is locked, immediately discard the
Monoject Magellan
Safety Needle and syringe into the nearest
sharps container following applicable regulations.
10. Puncture the skin with the One-Step™ Catheter assembly. A small
incision may be made at the puncture site with the # 11 Futura
®
Safety
Scalpel to ease catheter transition.
a. To Use Futura Safety Scalpel:
• Push slide forward with thumb until blade locks.
Retract blade before passing scalpel or disposing scalpel. To retract
apply pressure at the proximal end of the slide.
Scalpel is now ready for passing or disposal
11. When fluid flashback is observed, advance the catheter approximately
one centimeter and withdraw the needle completely.
WARNING: The needle must be withdrawn as soon as fluid is observed
to avoid puncturing internal organs.
12. The catheter may be repositioned during the procedure to facilitate
drainage, which also might involve moving of the patient to a more
convenient position for drainage.
13. Attach the three-way stopcock to the catheter. Laboratory samples may
be withdrawn at this time. Aspirate fluid with the 60ml syringe and inject
it into the 10ml specimen vials.
14. To drain fluid into a vacuum bottle, attach the tubing to the stopcock on
the catheter device. Close the clamp down completely and pierce the
vacuum bottle with the spike attached to the tubing.
15. Following the procedure, gently withdraw the catheter and dress the
site accordingly.
Monoject Magellan
is a trademark of Covidien AG
Manufacturer:
Merit Medical Systems, Inc.
South Jordan, Utah 84095 U.S.A. 1-801-253-1600
U.S.A. Customer Service 1-800-356-3748
www.merit.com
400641002/A ID 070810
SAFETY PARACENTESIS PROCEDURE TRAY
I n s t r u c t I o n s f o r u s e
PRODUCT DESCRIPTION:
The Safety Paracentesis Procedure Tray is a sterile, single use kit containing
devices designed to assist the clinician in draining fluid from body cavities.
INTENDED USE:
The Safety Paracentesis Procedure Tray is intended for drainage of fluid
and specimen collection for diagnostic purposes or to remove excess fluid
collections.
CONTRAINDICATIONS:
Thoracentesis
Pericardiocentesis
WARNINGS:
During catheter insertion procedure, if resistance is encountered, the cause
must be determined before continuing.
Withdraw the introducer needle as soon as fluid is observed to avoid
puncturing internal organs.
Refer to Lidocaine Hydrochloride Instructions for Use included in the box
for product use, contraindications, etc.
All sharp instruments (needles, scalpels) should be handled with care.
Activate the safety mechanism away from yourself or others.
To help avoid infectious disease due to accidental needlesticks,
contaminated needles should not be recapped or removed, unless there is
no alternative or such action is required by a medical procedure.
Used sharps are contaminated. Handle carefully. Dispose scalpel in
accordance with all applicable local and national laws and regulations.
CAUTIONS:
Read instructions prior to use.
RX Only.
Procedure must be performed by trained personnel.
Practitioners must be aware of complications associated with paracentesis.
Safety mechanism on the Monoject Magellan
Safety Needles
should not be deployed by placing hands in front of needle.
After deployment do not attempt to override or defeat the safety
mechanism on scalpel or needles.
This device is for single use only. Do not resterilize or reuse.
Store at controlled room temperature 15º to 30ºC (50º to 86ºF).
Use lidocaine ampules only if the solution is clear.
Do not Autoclave before use.
COMPLICATIONS / RISKS:
Perforation of organs Air embolism
Infection Hematoma
Pain in region Bleeding
Risks normally associated with percutaneous interventional procedures.
INSTRUCTIONS FOR USE:
Paracentesis is a sterile procedure which is typically performed with ultra-
sonic imaging, computed tomography imaging or fluoroscopic monitoring
and guidance. The procedure should be performed using aseptic technique.
1. It is recommended that the patient empty the bladder prior to
the procedure.
2. Open the paracentesis tray and unwrap the CSR wrap.
3. Prep the skin by using the skin cleansing agent.
4. Remove adhesive backing from the fenestrated drape and place it over the
puncture site. If the patient is in an upright position, the adhesive along the
edge of the drape may be used to secure upper portion.
Page view 0
1

Summary of Contents

Page 1

5. Aspirate the lidocaine by using the 18 gauge Monoject Magellan™ Safety Needle. Twist syringe clockwise while holding the rigid sheath that co